Improper Assembly-The AIRIS/AIRIS II MRI system has the potential for the input wiring of a power transformer to overheat due to damaged wiring or possible loose connections.
Source facts
Action
The recalling firm plans to inspect the transformer connections for damaged wiring and check the connections for tightness and tighten as necessary. The inspections and possible corrections began on 4/25/2006 and are in the process of being conducted by the firm''s Field Service Engineers. The firm anticipates trhat these actions will be completed by 6/25/2006.
Initiated
2006-04-25
Root cause
Other
Terminated
2008-10-29
Distribution
The devices were distributed to consignees located throughout the United States.
Recall number
Z-1115-06
Recall status
Terminated
CDRH record ID
45329
FDA classified
2006-06-14
Recall event number
35107