A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.
Source facts
Action
A field service engineer was dispatched to the customer site to perform the software update and to replace the PCB TSI-RSI board from 01/03/2024 through 12/26/2024.
Initiated
2024-01-03
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brazil, Croatia, France, Germany, Hungary, India, Ireland, Italy, New Zealand, Portugal, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland.
Recall number
Z-1115-2025
Recall status
Completed
CDRH record ID
212003
FDA classified
2025-02-06
Recall event number
96103