Skip to content

Public FDA record

Z-1116-2012: Perkin Elmer Genetic Screening Processor (GSP) Product code: 2021-0010. It is intended for in vitro quant

Recall record Z-1116-2012 for PerkinElmer Life and Analytical Sciences, Wallac, OY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1116-2012
Company in FDA record
PerkinElmer Life and Analytical Sciences, Wallac, OY
FDA classification date
Feb 28, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

GSP instruments with defective washer modules may cause a false negative screening result being reported for T4, TSH and IRT. For 17-OHP, a false positive result may be generated if the failure occurs for a normal sample.

Source facts

Action
Perkin Elmer issued an "Urgent Medical Device Recall" letter dated 1/25/12 to all customers and informed them about the possible issue and requested permission for service personnel to check their instrument. The customers were instructed that the instrument could not be used and the washer module required replacement. Customers were told to expect to be contacted by the firm's local representative for further details.
Initiated
2012-01-09
Root cause
Nonconforming Material/Component
Terminated
2013-03-06
Distribution
Worldwide Distribution - USA, including the states of Florida, Ohio, Oregon, Texas and the countries of Brazil, Belgium, France, Germany, and the UK.
Recall number
Z-1116-2012
Recall status
Terminated
CDRH record ID
107141
FDA classified
2012-02-28
Recall event number
61031

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗