PerkinElmer Life and Analytical Sciences, Wallac, OY
FDA classification date
Feb 28, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
GSP instruments with defective washer modules may cause a false negative screening result being reported for T4, TSH and IRT. For 17-OHP, a false positive result may be generated if the failure occurs for a normal sample.
Source facts
Action
Perkin Elmer issued an "Urgent Medical Device Recall" letter dated 1/25/12 to all customers and informed them about the possible issue and requested permission for service personnel to check their instrument. The customers were instructed that the instrument could not be used and the washer module
required replacement. Customers were told to expect to be contacted by the firm's local representative for further details.
Initiated
2012-01-09
Root cause
Nonconforming Material/Component
Terminated
2013-03-06
Distribution
Worldwide Distribution - USA, including the states of Florida, Ohio, Oregon, Texas and the countries of Brazil, Belgium, France, Germany, and the UK.
Recall number
Z-1116-2012
Recall status
Terminated
CDRH record ID
107141
FDA classified
2012-02-28
Recall event number
61031