Skip to content

Public FDA record

Z-1116-2022: Olympus Uretero-reno videoscope, Model No. URF-V2

Recall record Z-1116-2022 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1116-2022
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
May 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.

Source facts

Action
On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.
Initiated
2022-03-25
Root cause
Process change control
Distribution
US Nationwide distribution in the state of AZ.
Recall number
Z-1116-2022
Recall status
Open, Classified
CDRH record ID
193265
FDA record created
2022-05-19
Recall event number
90063

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗