Skip to content

Public FDA record

Z-1116-2025: Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in conjunction with the SmartD

Recall record Z-1116-2025 for Max Mobility LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1116-2025
Company in FDA record
Max Mobility LLC
FDA classification date
Feb 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to faulty circuit board in speed control remote use in conjunction with a wheelchair power assist, their is a potential for lost of control of device potentially resulting in minor and major injuries.

Source facts

Action
Initially Max Mobility notified consignees of their Urgent Medical Correction between the dates of 12/20/24 - 01/17/2025. Since this time, Max Mobility has expanded the recall due to recalled product that may have been inadvertently comingled with redesigned product and shipped after December 20, 2024. Starting 3/25/25, Max Mobility began reaching out to customer via phone, email and mail about the expansion. In addition, Max Mobility updated their press release on 4/7/2025 1. Click on the link or scan QR code to enter the portal within the letter 2. Complete requested portal information. 3. If you are a dealer and have distributed the affected product to someone else, you will be asked to select a transmission method. You may either (a) notify your end users directly and include a copy of this notice with your communication or (b) provide end user contact information to Max Mobility through the online portal and we will contact them for you. 4. After acknowledgement, instructions for obtaining replacement SpeedControl Dial units will be provided. You will need the associated client ID provided in ATTACHMENT 1 of this letter when requesting SpeedControl Dial replacement(s). 5. Once the replacement SpeedControl Dial has been issued to the end user, you will be asked to verify that the correction was completed via the field action portal and destroy or return the effected SpeedControl Dial to Max Mobility.
Initiated
2024-12-20
Root cause
Component change control
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Sweden, Canada, Australia, Japan, New Zealand, Colombia, Argentina, Hong Kong, Canada, Mexico, Bermuda, South Korea, Chile
Recall number
Z-1116-2025
Recall status
Open, Classified
CDRH record ID
211972
FDA classified
2025-02-11
Recall event number
96086

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗