Z-1117-2020: Rusch Greenlite Single-Use Metal Laryngoscope Blade for Fiber Optic System

Recall record Z-1117-2020 for TELEFLEX MEDICAL INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1117-2020
Company in FDA record
TELEFLEX MEDICAL INC
FDA classification date
Feb 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The products are being recalled due to reports of breakage broke at the point of a welded joint either prior to use during setup or during use inside the patient's mouth.

Source facts

Action
Letters sent to customers/users to immediately discontinue use and quarantine any products with the affected product code and lot numbers and return them to the manufacturer.
Initiated
2020-01-09
Root cause
Process control
Terminated
2023-09-11
Distribution
Distributed nationwide in the US and to Puerto Rico; OUS to more than 20 countries; To the following government VA medical Centers: VAMC - KANSAS CITY MO 4801 E Linwood Blvd Kansas City MO 64128-2218
Recall number
Z-1117-2020
Recall status
Terminated
CDRH record ID
179079
FDA classified
2020-02-07
Recall event number
84702

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗