Z-1118-2025: UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929

Recall record Z-1118-2025 for Medicrea International. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1118-2025
Company in FDA record
Medicrea International
FDA classification date
Feb 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.

Source facts

Action
On 12/03/2024, Medtronic sent an "Urgent Medical Device Correction" notification to physician users of the UNiD HUB informing them of a software anomalies in the UNiD" HUB that may have impacted UNiD rod planning and certain optional surgical parameters (Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification), and may have errors that resulting in incorrect calculations displayed in the UNiD" HUB. Medtronic is also informing users of a Software update that has removed the impacted parameters. The software update was released on 12/03/2024. Customer are informed of Patient Management Recommendations: If a UNiD rod was planned without any of these four parameters and met the surgeons planned surgical objectives, no additional actions are required. As these parameters are optional for planning, there is no record of whether the parameters were utilized for surgical planning. If you are concerned that the use of these parameters may have adversely impacted a patient s sagittal balance, it is recommended to re-evaluate the surgical plan and continue to follow the patient based on your assessment of their post operative sagittal balance. Monitor patients per routine standard of care and intervene per medical judgement. Customer are instructed to: " Review this information within this letter " Healthcare providers may continue to use the UNiD HUB as instructed in the External User Guide. " Confirm via the enclosed confirmation form that this notification has been communicated within their facility with all physician users. Send the completed Customer Confirmation Form to Medtronic viaemail at neuro.quality@medtronic.com " This notice needs to be passed on to those who need to be aware within their organization that utilizes the UNiD HUB for planning and to maintain a copy of this notice in their records.
Initiated
2024-12-03
Root cause
Software design
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.: Argentina Australia Belgium Brazil Bulgaria Chile Czech Republic Denmark Finland France Germany Greece Iceland Ireland Israel Italy Luxembourg Malaysia Mexico Monaco Morocco Netherlands New Zealand Poland Portugal Singapore South Korea Spain Sweden Switzerland United Kingdom
Recall number
Z-1118-2025
Recall status
Open, Classified
CDRH record ID
211640
FDA classified
2025-02-07
Recall event number
95957

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗