Skip to content

Public FDA record

Z-1119-2020: LivaNova Sorin Connect DPIII KOMPL. S5 REF 96-415-130, 2017-08-15, IPX1, RXOnly, (01)04033817901594 Prod

Recall record Z-1119-2020 for LivaNova USA Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1119-2020
Company in FDA record
LivaNova USA Inc.
FDA record creation date
Feb 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Incorrect labeling; it was identified that some Datapad devices contained incorrect labels. Specifically the manufacturing date in the production identifier was incorrect and the manufacturing date on the unique device identifier label was labeled in the format of yy-mm-dd instead of yyyy-mm-dd.

Source facts

Action
On 01/28/2019, a Medical Device Correction letter was mailed to affected consignees. The letter asked customers to take the following actions: 1. Please check your inventory using the list of affected products delivered to your facility in Attachment 1; 2. Using the attached Customer Response Form (Attachment 2), please respond by email to USFSN@livanova.com to confirm that you have affected products in your facility. Your LivaNova reference person will contact you for the replacement of the product labels with labels bearing the correct manufacturing date. 3. Please ensure that this Medical Device Correction is communicated to all personnel within your organization who need to be aware of it. If you have transferred any of the affected devices to a third party, please communicate this information to them and inform the LivaNova Quality Assurance Team at USFSN@livanova.com. 4. For questions regarding this Medical Device Correction, please contact us at (800) 986-4702 or by e-mail to USFSN@livanova.com.
Initiated
2019-01-28
Root cause
Error in labeling
Terminated
2022-01-31
Distribution
US: NJ and CA
Recall number
Z-1119-2020
Recall status
Terminated
CDRH record ID
177393
FDA record created
2020-02-10
Recall event number
84142

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗