Skip to content

Public FDA record

Z-1119-2023: Surgical convenience kits labeled as: a) CENTURION MEDICAL PRODUCTS CHEST TUBE INSERTION TRAY, Reorder N

Recall record Z-1119-2023 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1119-2023
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Feb 10, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.

Source facts

Action
The firm sent a correction notice to its direct consignees on 12/07/2022 by US mail and email. The notice explained the issue and requested the following: "Please check you stock for affected kits and over-labeling will be sent to place on kits to ensure that all labeling is accurate. No kits are to be returned back to Medline. All components are not affected. Our records indicate that these were used as a component in the list of kit(s) purchased from Medline Industries, LP." "If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP." "If you have any questions, please contact the Recall Department at 866-359-1704."
Initiated
2022-12-07
Root cause
Under Investigation by firm
Distribution
Nationwide
Recall number
Z-1119-2023
Recall status
Open, Classified
CDRH record ID
198021
FDA classified
2023-02-10
Recall event number
91483

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗