Z-1119-2025: (1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis

Recall record Z-1119-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1119-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Feb 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software issues could potentially result in: 1) delays in accessing dispensing software application, 2) wrong therapy (dose, item, timing, etc.), or 3) unauthorized access and/or compliance related issues

Source facts

Action
On January 8, 2025, the firm notified customers via email and mailed letter titled "Urgent Medical Device Correction". Firm states customers can continue to use Pyxis products in accordance with their labeling and any previously provided field action notifications, including the following guidance: 1. Facilities should establish policies and procedures for when an automated dispensing system is unavailable. 2. Device Keys can manually access unavailable drawers, but must be readily available and retrievable to adequately mitigate the risk. 3. Emergency supplies should be physically available in care areas for critical or urgent medications. 4. If using critical override mode to access contents of the device, users must use extra precautions and verify correct medications are selected. The firm anticipates releasing software update ES 1.11 beginning June 2025, which will address these software issues.
Initiated
2025-01-08
Root cause
Software design
Distribution
Domestic distribution nationwide. International distribution to North America, Latin America, EMEA, and Greater Asia.
Recall number
Z-1119-2025
Recall status
Open, Classified
CDRH record ID
211880
FDA classified
2025-02-07
Recall event number
96055

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗