Difficulty in pressing the button needed to release a bed part resulted in an injury to a caregiver's finger which became stuck in metal latch under the gas cylinder of the bed piece.
Source facts
Action
An official notification letter dated February 3, 2009 was hand-carried by a local Berchtold Corporation Technical Service representative to each account. The letter indicated that the Berchtold Corporation Service Technician will complete recommended upgrades. Direct questions concerning this recall to your BERCHTOLD Technician Service at 1-800-243-5135 and press the Number 2.
Initiated
2009-02-02
Root cause
Device Design
Terminated
2009-06-04
Distribution
Worldwide Distribution -- US and Canada.
Recall number
Z-1123-2009
Recall status
Terminated
CDRH record ID
79572
FDA classified
2009-04-27
Recall event number
51276