Skip to content

Public FDA record

Z-1123-2021: Gambro Cartridge Low Weight-Low Volume Blood Transport System for Hemodialysis Blood Set, product code 11

Recall record Z-1123-2021 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1123-2021
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Feb 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

complaints of kinked access lines observed during treatment using cartridge sets for hemodialysis treatments were received.

Source facts

Action
The firm issued a Safety Alert on 12/30/2020 by mail. The letter explained the problem with photographs, and the hazard involved with using a device with a kinked line. Customers identifying set with kinked tubing may contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Distributor were asked to notify their customers by distributing the letter.
Initiated
2020-12-30
Root cause
Under Investigation by firm
Distribution
Worldwide distribution.
Recall number
Z-1123-2021
Recall status
Open, Classified
CDRH record ID
185444
FDA record created
2021-02-16
Recall event number
87150

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗