Skip to content

Public FDA record

Z-1124-03: Intralase FS Laser System. Laser Keratome

Recall record Z-1124-03 for Intralase Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1124-03
Company in FDA record
Intralase Corp
FDA classification date
Aug 12, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Gantry moves toward patient when 'Home' button is used after procedure.

Source facts

Action
Firm sent Safety Alert letters to customers on July 15, 2003 by registered mail instructing not to use the HOME button, including a 'Do Not Use' sticker for the machine and an addendum to the Operations Manual.
Initiated
2003-07-15
Root cause
Other
Terminated
2004-12-02
Distribution
Nationwide and Japan.
Recall number
Z-1124-03
Recall status
Terminated
CDRH record ID
28521
FDA classified
2003-08-12
Recall event number
26826

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗