Open original FDA record ↗Recall record
- FDA identifier
- Z-1124-2013
- Company in FDA record
- Biomet U.K., Ltd.
- FDA classification date
- Apr 15, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
Source facts
- Action
- BIOMET sent an Urgent Medical Device Recall Notice on March 20, 2013, to all affected customers with directions to forward the notice if product was further distributed to Hospital Operating Room managers. The notice identified the recalled device, reason for the recall, and possible adverse events directing consignees to locate and discontinue use of the device. A FAX back Response form was to be returned to 574-372-1683 within 3 days of receipt prior to return of the product to Biomet, USA. Questionsregarding this recall should be addressed to 574 372-1570 M-F, 8a.m. -5 p.m.
- Initiated
- 2013-03-20
- Root cause
- Device Design
- Terminated
- 2015-05-04
- Distribution
- Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
- Recall number
- Z-1124-2013
- Recall status
- Terminated
- CDRH record ID
- 116755
- FDA classified
- 2013-04-15
- Recall event number
- 64679