Reason stated in the source
COBAS INTEGRA 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
Source facts
- Action
- Roche Diagnostics issued on 12/14/18 "Urgent Medical Device Correction (UMDC) via UPS Ground (receipt required) Consignee Instructions: Manually add the EWCs listed in the Urgent Medical Device Correction (UMDC) on your COBAS INTEGRA 400 plus and cobas c 111 analyzers. If customers run the test combinations requiring needle assignment (See Table on page 3 of UMDC TP-00508), run those in batch mode until the updated TAS has been loaded on the COBAS INTEGRA 400 Plus analyzer. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions.
- Initiated
- 2018-12-14
- Root cause
- Device Design
- Terminated
- 2020-09-04
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1124-2019
- Recall status
- Terminated
- CDRH record ID
- 169985
- FDA record created
- 2019-04-10
- Recall event number
- 81814