Z-1124-2021: Getinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector intended for

Recall record Z-1124-2021 for Getinge Group Logistics America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1124-2021
Company in FDA record
Getinge Group Logistics America, LLC
FDA record creation date
Feb 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.

Source facts

Action
On 01/12/2021 Getinge initiated a voluntary Medical Device in the U.S for the CM320 Series Washer Disinfector instructing consignees to inspect their device and that a correction kit is currently available and the correction will be completed at no cost to affected customers.
Initiated
2020-12-21
Root cause
Device Design
Terminated
2022-07-20
Distribution
Worldwide distribution - US Nationwide distribution including in the states of ND, PA, NC, VA, NE, MD, NY, MN and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Norway, Slovakia, Slovenia, South Korea, Sweden, Japan, China, Netherland, Italy, Kuwait and Czech Republic.
Recall number
Z-1124-2021
Recall status
Terminated
CDRH record ID
185452
FDA record created
2021-02-16
Recall event number
87155

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗