Open original FDA record ↗Recall record
- FDA identifier
- Z-1124-2021
- Company in FDA record
- Getinge Group Logistics America, LLC
- FDA record creation date
- Feb 16, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.
Source facts
- Action
- On 01/12/2021 Getinge initiated a voluntary Medical Device in the U.S for the CM320 Series Washer Disinfector instructing consignees to inspect their device and that a correction kit is currently available and the correction will be completed at no cost to affected customers.
- Initiated
- 2020-12-21
- Root cause
- Device Design
- Terminated
- 2022-07-20
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of ND, PA, NC, VA, NE, MD, NY, MN and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Norway, Slovakia, Slovenia, South Korea, Sweden, Japan, China, Netherland, Italy, Kuwait and Czech Republic.
- Recall number
- Z-1124-2021
- Recall status
- Terminated
- CDRH record ID
- 185452
- FDA record created
- 2021-02-16
- Recall event number
- 87155