Open original FDA record ↗Recall record
- FDA identifier
- Z-1124-2022
- Company in FDA record
- Devicor Medical Products Inc
- FDA record creation date
- May 23, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Small patient labels do not match the primary device label. The primary device label accurately identifies the product as 4010-05-08-T4, while the smaller patient label incorrectly identifies the product as 4010-05-08-T3.
Source facts
- Action
- On March 30, 2022, the firm distributed Urgent Medical Device Recall letters to their customers via email. Customers were informed of the labeling issue. Users were instructed to examine their inventory and identify any affected product. Users should contact their local sales representative for correction of any affected devices. The sales representative will be on site to remove the affected patient labels. Please do not use affected patient labels from this lot. If you have any additional questions, comments or concerns please contact the firm's Customer Support at +1 877-926-2666. You may also email mmtchnrecall@mammotome.com.
- Initiated
- 2022-03-30
- Root cause
- Labeling Change Control
- Terminated
- 2024-04-30
- Distribution
- US Nationwide distribution in the states of DE, KY, NY, OH, WV.
- Recall number
- Z-1124-2022
- Recall status
- Terminated
- CDRH record ID
- 193511
- FDA record created
- 2022-05-23
- Recall event number
- 90119