Open original FDA record ↗Recall record
- FDA identifier
- Z-1126-2013
- Company in FDA record
- Accuray Incorporated
- FDA classification date
- Apr 16, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.
Source facts
- Action
- On 3/19/13, an "Urgent Device Correction Letter" was distributed to all affected users, with a description of the problem and user corrective action steps. US customers can contact 1-877- 668-8667 if questions.
- Initiated
- 2013-03-19
- Root cause
- Component design/selection
- Terminated
- 2013-10-23
- Distribution
- Worldwide distribution, including Nationwide in the US and the countries of Korea, Myanmar, Canada, Belgium, England, Czech Republic, Finland, France, Germany, Greece, India, Italy, Poland, Russia, Spain, Switzerland, Turkey, Columbia, Mexico, Japan and Saudi Arabia.
- Recall number
- Z-1126-2013
- Recall status
- Terminated
- CDRH record ID
- 116839
- FDA classified
- 2013-04-16
- Recall event number
- 64712