Z-1126-2015: BodyGuard 323 pump, models 100-510PXSI, 100-516PXS, 100-517PXS, 100-518PXS, 100-603XSA and 100-603XSAP

Recall record Z-1126-2015 for CME America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1126-2015
Company in FDA record
CME America, LLC
FDA classification date
Feb 19, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

CME America is recalling the BodyGuard and BodyGuard 323 Infusion pumps due to the potential for an over delivery.

Source facts

Action
CME America sent an "Urgent Medical Device Recall" letter dated January 16, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Required Action for Users: To date there have been no reports of adverse events or serious injury. However, there is the potential for serious injury or death. CME America recommends that you DO NOT use the BodyGuard 323 pump for PCA infusion. If your devices are currently in use, be sure that the programmed parameters DO NOT involve a patient controlled bolus dose less than 0.1 miln volume (translated into meg or mg). Follow-up Actions by CME America: CME America has identified the root cause of this issue and is recommending installation of a new version of software, USA36B4, for affected BodyGuard 323 infusion pumps. This software is identical to the previously released version, USA36A4.2. However the designation of B4 in the software name will identify updated software in each device. CME America will contact all affected customers within 60 days to initiate the scheduling process for the remediation. Your assistance is appreciated and necessary to prevent use and facilitate remediation of the recalled product. CME America can address any questions regarding the recalled product by calling the CME America recall line at 720-839-2391.
Initiated
2015-01-16
Root cause
Software design
Terminated
2017-03-31
Distribution
Worldwide Distribution - Nationwide Distribution and VA/military/govt consignees and the country of Canada.
Recall number
Z-1126-2015
Recall status
Terminated
CDRH record ID
132890
FDA classified
2015-02-19
Recall event number
70294

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗