Reason stated in the source

Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.

Source facts

Action
On March 13 , 2019 the firm sent letters to their consignees stating the following: The purpose of this letter is to advise you that ulrich medical USA is voluntarily recalling certain Torque Limiting Handles. Reason for the voluntary recall: Certain Torque Limiting Handles in the field are beyond their 3 year calibration life. The calibration issue experienced to date has been limited to select instruments and not all of these instruments have been affected. Due to the nature of how these instruments are used the risk to the patient is low. No injury or death has been reported. Risk to health: Over or under tightening of the set screw could cause the construct to fatigue or loosen. The risk to the patient is low. Actions to be taken by the user: Return the torque limiting handles to ulrich medical USA at ulrichs expense, and a new torque limiting handle will be shipped back. Instructions for acknowledgement of receipt include FedEx signature required. Actions to be taken by the Company: All affected instruments will be replaced with new torque limiting handle upon return. Contact information: If you have any questions or concerns please contact: Louis Milos Sr. Manager Business and Quality Systems 18221 Edison Ave. Chesterfield, MO 63005 636-519-0268 ext. 105 l.milos@ulrichmedicalusa.com www.ulrichmedicalusa.com Available Monday through Friday 8:00am to 5:00pm CST
Initiated
2019-03-08
Root cause
Storage
Terminated
2020-07-10
Distribution
AZ, IL, LA, MI, MO, NV, OH, PA, TN, and TX.
Recall number
Z-1126-2019
Recall status
Terminated
CDRH record ID
171706
FDA record created
2019-04-11
Recall event number
82497

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗