Skip to content

Public FDA record

Z-1126-2021: Orient Gene COVID- 19 IgG/ IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) REF GCCOV-402a - Product Us

Recall record Z-1126-2021 for Confirm Biosciences Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1126-2021
Company in FDA record
Confirm Biosciences Inc
FDA record creation date
Feb 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to two (2) issues: 1) The firm distributed COVID-19 Rapid Test Cassettes outside of the Emergency Use Authorization to non-certified Clinical Laboratories. 2) COVID-19 Rapid Test Cassettes were not adequately validated for its intended use (fingerstick tests for in-home use/testing).

Source facts

Action
On 01/29/2021, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Notification its customers and distributors informing them that Recalling Firm may have distributed products to facilities that do not meeting the conditions outline in the Emergency Use Authorization (EUA) which stipulates that only laboratories certified under Clinical Laboratory Improvement Amendments of 1988 (CLIA) to perform moderate or high complexity tests are authorized to receive and utilize the products. Use of these tests by non-certified laboratory facilities may result in false-positive test results and may lead to an incorrect assessment, such as mistaken immune response to the COVID-19 virus and may delay treatment and risk the further spread of infection to others. The Recalling Firm is planning on sending a follow-up Customer Notification Letter via UPS on or about 02/12/2021. The "URGENT: MEDICAL DEVICE RECALL" Notification instructs customers: 1) To minimize unauthorized use of these products, discontinue the use these products outside of a CLIA accredited or compliant Moderate or High Complexity laboratory setting. 2) Acknowledge the receipt of this notice by completing the attached Response Form, scanning and emailing the completed form to: Regulatory@confirmbiosciences.com Customer are asked to complete the response form and return immediately, no later than March 31, 2021. To correct this issue, the Recalling Firm will work with its customers to either: -Affirm the Customers authorization to use these products through their CLIA Number and Status -Coordinate the return of unused, unopened products to the Recalling Firm. For any questions regarding this recall, call 800-908-5602 or email Regulatory@confirmbiosciences.com (Monday thru Friday 8:00 a.m. to 4:00 p.m. Pacific Time)
Initiated
2021-01-29
Root cause
Other
Terminated
2022-12-13
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. The countries of Australia, Mexico, Honduras, Spain, Sierra Leone, Bahamas, Ecuador, Turkey, Romania, Kuwait, Libya, Qatar, Ghana, Argentina, Bolivia, Brazil, Guatemala, Panama, India, United Kingdom, West Indies, South Africa, United Arab Emirates, Slovenia, Canada, Israel, United Kingdom, Romania, Argentina, Honduras, Russian Federation, R¿publique D¿mocratique du Congo, Switzerland, Dominican Republic, Greece, Saudi Arabia, and Portugal.
Recall number
Z-1126-2021
Recall status
Terminated
CDRH record ID
185665
FDA record created
2021-02-16
Recall event number
87263

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗