ACL TOP unit, when in emergency stop may release liquid during recovery sequence and contaminate sample tube.
Source facts
Action
Instrumentation Laboratory notiifed the distributor Beckman Coulter on 6/19/2007. Beckman Coulter notified their customers and provided a copy with the Instructions for Recovery Procedure on how to avoid contamination from probe leaks.
Initiated
2007-06-19
Root cause
Other
Terminated
2012-05-23
Distribution
Nationwide.
Recall number
Z-1127-2007
Recall status
Terminated
CDRH record ID
53218
FDA classified
2007-08-07
Recall event number
38234