Z-1127-2007: Instrumentation Laboratory ACL TOP Coagulation Analyzer (Base Model only). Blood analyzer. P/N 280000

Recall record Z-1127-2007 for Instrumentation Laboratory Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1127-2007
Company in FDA record
Instrumentation Laboratory Co.
FDA classification date
Aug 7, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

ACL TOP unit, when in emergency stop may release liquid during recovery sequence and contaminate sample tube.

Source facts

Action
Instrumentation Laboratory notiifed the distributor Beckman Coulter on 6/19/2007. Beckman Coulter notified their customers and provided a copy with the Instructions for Recovery Procedure on how to avoid contamination from probe leaks.
Initiated
2007-06-19
Root cause
Other
Terminated
2012-05-23
Distribution
Nationwide.
Recall number
Z-1127-2007
Recall status
Terminated
CDRH record ID
53218
FDA classified
2007-08-07
Recall event number
38234

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗