Skip to content

Public FDA record

Z-1127-2021: AntIbody to SARS-CoV-2 - Product Usage: (Colloidal Gold) for IgG/IgM Diagnostic Kit, RAPID TEST DEVICE, W

Recall record Z-1127-2021 for Sky Medical Supplies and Equipments LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1127-2021
Company in FDA record
Sky Medical Supplies and Equipments LLC
FDA record creation date
Feb 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Distributed COVID test kits without emergency use authorization.

Source facts

Action
Sky Medical initiated the recall on 01/19/2021 by telephone. The firm sent a follow-up letter disseminated by email on 02/03/2021. The firm is requesting the return of the product.
Initiated
2021-01-19
Root cause
No Marketing Application
Terminated
2026-03-10
Distribution
Worldwide distribution - US Nationwide distribution in the state of Florida and the country of Peru.
Recall number
Z-1127-2021
Recall status
Terminated
CDRH record ID
185763
FDA record created
2021-02-16
Recall event number
87300

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗