Open original FDA record ↗Recall record
- FDA identifier
- Z-1128-2012
- Company in FDA record
- Zeiss, Carl Inc
- FDA classification date
- Feb 29, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The manufacturer confirmed that the opto-semiconductor sensors can be triggered by exposure to incandescent lighting. The safety interlock is in place to prevent unintended exposure to Class 3B laser radiation.
Source facts
- Action
- Carl Zeiss MicroImaging LLC, through trained personnel, first visited each user site, provided additional user instructions, re-labeled the components to provide adequate instruction and warning and further placed a label over the access to the locking screw which secures the component to the microscope. Carl Zeiss Inc. issued "IMPORTANT LASER SAFETY NOTICE: URGENT: FIELD CORRECTIVE ACTION" notices to each user by express courier in late April 2011, Carl Zeiss personnel will confirm that the relabeling has been performed and updated instructions for use have been provided. For questions, please contact the company. The firm will implement 2 actions. Service Representatives will visit each user site, provide additional user instructions, relabel the components to provide adequate instruction and warning, and add a label over the access to the locking screw which secures the component to the microscope. Secondly, when new replacement sensors are available, Carl Zeiss field service technicians will perform the component replacements. Modified components are anticipated to be available within 6 months and replacement completed by April 2012.
- Initiated
- 2011-04-28
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-12-11
- Distribution
- Worldwide Distribution -- USA (nationwide)
- Recall number
- Z-1128-2012
- Recall status
- Terminated
- CDRH record ID
- 102128
- FDA classified
- 2012-02-29
- Recall event number
- 59285