SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.
Source facts
Action
SR Instruments, Inc. sent a Notification letter dated December 13, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions call 716-693-5977.
Initiated
2012-12-13
Root cause
Use error
Terminated
2015-03-12
Distribution
USA Nationwide Distribution in the state of Texas.
Recall number
Z-1128-2013
Recall status
Terminated
CDRH record ID
116806
FDA classified
2013-04-17
Recall event number
64698