Open original FDA record ↗Recall record
- FDA identifier
- Z-1128-2015
- Company in FDA record
- Mako Surgical Corporation
- FDA classification date
- Feb 19, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activated
Source facts
- Action
- On January 22, 2015 the firm sent an IMPORTANT MEDICAL DEVICE CORRECTION asking consignees to complete the attached acknowledgement form and return them by email, fax or mail. The notice also noted the RIO system can continue to be used. The likelihood of occurrence is rare (.035% of cases) and the event is readily detectable and can be resolved. Should you have any questions regarding this Important Medical Device Recall Notice, to contact us at (954) 628-1721.
- Initiated
- 2015-01-22
- Root cause
- Device Design
- Terminated
- 2016-12-14
- Distribution
- Nationwide Distribution-including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii.
- Recall number
- Z-1128-2015
- Recall status
- Terminated
- CDRH record ID
- 133374
- FDA classified
- 2015-02-19
- Recall event number
- 70374