Skip to content

Public FDA record

Z-1129-2021: System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition

Recall record Z-1129-2021 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1129-2021
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Feb 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.

Source facts

Action
The manufacturer sent an "Urgent Medical Device Correction" letter (GEHC Ref# 25496) to customers with affected systems on 12/17/2020.
Initiated
2020-12-17
Root cause
Process change control
Terminated
2023-01-09
Distribution
Worldwide - US Nationwide Distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NV, NY, NH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV; In the countries of Algeria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Denmark, Estonia, Finland, France, Gabon, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea (Republic Of), Kuwait, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uzbekistan, Viet Nam.
Recall number
Z-1129-2021
Recall status
Terminated
CDRH record ID
185314
FDA record created
2021-02-17
Recall event number
87089

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗