Skip to content

Public FDA record

Z-1129-2025: Cardinal Health Presource Kits: 1) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV20110B; 2) KIT, CAROTID

Recall record Z-1129-2025 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1129-2025
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Feb 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.
Initiated
2024-11-08
Root cause
Material/Component Contamination
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.
Recall number
Z-1129-2025
Recall status
Open, Classified
CDRH record ID
211604
FDA classified
2025-02-10
Recall event number
95947

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗