Open original FDA record ↗Recall record
- FDA identifier
- Z-1130-03
- Company in FDA record
- Abbott Laboratories HPD/ADD
- FDA classification date
- Aug 13, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Free T4 Controls, may yield control values that are high and outside of the package insert ranges.
Source facts
- Action
- The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, were recalled by letter dated 5/30/03. The accounts were informed of the out of range control values, and the potential for depressed patient values with the IMx kit, and were requested to discontinue use and destroy the affected lots of product. Abbott sent a Device Correction letter dated 6/4/03 to those accounts receiving the Architect Free T4 Calibrator kits, informing them of the problems experienced with the The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, made from the same raw material. The accounts were informed that the Architect Calibrator was performing as expected, but as a precaution recommended conducting a daily QC with non-Abbott Controls and to report back to Abbott any out of range non-Abbott control results.
- Initiated
- 2003-05-30
- Root cause
- Other
- Terminated
- 2003-09-16
- Distribution
- Nationwide and internationally to Abbott subsidiaries in Ecuador, Uruguay, Argentina, Brazil, El Salvador, Canada, Germany, Japan, Singapore, Hong Kong, Thailand, Korea, Australia, Costa Rica, Grenada, St. Lucia, Bermuda, Antigua, Panama, Guatemala, Mexico, Colombia, Jamaica, Paraguay, Peru, Chile, Dominican Republic, United Kingdom, New Zealand, Trinidad & Tabago, St. Vincent, Barbados, Curacao, St. Kitts, Anguilla, Turks & Caicos Islands, Cayman Islands, Germany and Taiwan.
- Recall number
- Z-1130-03
- Recall status
- Terminated
- CDRH record ID
- 27666
- FDA classified
- 2003-08-13
- Recall event number
- 26430