Skip to content

Public FDA record

Z-1130-2022: Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2V

Recall record Z-1130-2022 for Invacare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1130-2022
Company in FDA record
Invacare Corporation
FDA record creation date
May 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery harness results in loss of power to the wheelchair.

Source facts

Action
A preliminary notification was sent to providers on March 30, 2022, introducing the product issue and providing instructions regarding how to handle impacted units before field correction kits are available. A notification letter dated April 13, 2022, was sent to impacted consignees via traceable mail. This mailing includes a Provider Letter which will introduces the field correction requirements and provides instructions for acknowledging awareness and understanding of the requested field correction activities. Providers should inform affected customers to inform them that their wheelchair is subject to recall and requires repair. The letter provides user precautions that must be taken until the wheelchair is repaired. If you have any questions concerning this recall, please call Invacare s field correction support line at (877) 413-6008
Initiated
2022-03-30
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AR, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, and WI. The country of Canada.
Recall number
Z-1130-2022
Recall status
Open, Classified
CDRH record ID
193354
FDA record created
2022-05-25
Recall event number
90084

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗