Skip to content

Public FDA record

Z-1131-2021: Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder

Recall record Z-1131-2021 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1131-2021
Company in FDA record
Biomet, Inc.
FDA record creation date
Feb 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.

Source facts

Action
On 01/19/2019, Zimmer Biomet issued Urgent Medical Device Recall notices to distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory will be notified via FedEx. Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
Initiated
2021-01-15
Root cause
Labeling Change Control
Terminated
2023-03-09
Distribution
US Nationwide distribution including in the states of MS, MN, KY, FL, MI, OH, AZ, IL, WI, CA, UT, TN.
Recall number
Z-1131-2021
Recall status
Terminated
CDRH record ID
185528
FDA record created
2021-02-17
Recall event number
87201

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗