Open original FDA record ↗Recall record
- FDA identifier
- Z-1131-2021
- Company in FDA record
- Biomet, Inc.
- FDA record creation date
- Feb 17, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
Source facts
- Action
- On 01/19/2019, Zimmer Biomet issued Urgent Medical Device Recall notices to distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory will be notified via FedEx. Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
- Initiated
- 2021-01-15
- Root cause
- Labeling Change Control
- Terminated
- 2023-03-09
- Distribution
- US Nationwide distribution including in the states of MS, MN, KY, FL, MI, OH, AZ, IL, WI, CA, UT, TN.
- Recall number
- Z-1131-2021
- Recall status
- Terminated
- CDRH record ID
- 185528
- FDA record created
- 2021-02-17
- Recall event number
- 87201