Open original FDA record ↗Recall record
- FDA identifier
- Z-1132-2009
- Company in FDA record
- Medtronic Neurosurgery
- FDA classification date
- Mar 27, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The ventricular catheters may become detached from the Snap Base assembly after implantation and this condition may increase the need for emergency revision surgery.
Source facts
- Action
- An Urgent - Medical Device Recall notification was sent on February 13,2009 to the attention of risk managers, advising them to cease use of the Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782 ( all versions and all lot numbers), the Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27802 ( all versions and all lot numbers), and the Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27708 ( all version and all lot numbers). Each customer is to cease the use of the listed product and any unused unit needs to be returned to Medtronic. Further, each account is required to account for product used/implanted and to advise surgeons of the recall information. Each account will recieved an individualized Customer Product Reconciliation Form provided by their Medtronic Sales Representative for determining the status of any product shipped to them.
- Initiated
- 2009-02-12
- Root cause
- Device Design
- Terminated
- 2010-11-08
- Distribution
- Worldwide Distribution -- USA, Ireland, and Australia.
- Recall number
- Z-1132-2009
- Recall status
- Terminated
- CDRH record ID
- 79642
- FDA classified
- 2009-03-27
- Recall event number
- 51287