Z-1132-2018: NxStage PureFlow B Solution - Premixed Dialysate for Hemodialysis with NxStage System One

Recall record Z-1132-2018 for NxStage Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1132-2018
Company in FDA record
NxStage Medical, Inc.
FDA record creation date
Mar 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Using different fluid formulations of NxStage PureFlow B Solution at the same time during treatment may cause patient health risk. The NxStage PureFlow B Solution is for use with systems that use one premixed dialysate during hemodialysis. The original labeling did not warn against mixing dialysate formulations and did not instruct staff to verify that the fluid is flowing during use.

Source facts

Action
Customers are advised to : 1. Always make sure when using PureFlow B Solution that fluid is flowing freely from all bags. 2. Recognize that any practice of hanging more than one fluid type at the same time is not an intended use of the NxStage System One" (as it does not monitor fluid composition) and is not a labeled indication of PureFlow B solutions. 3. Please make sure that all users of PureFlow B Solution in your organization and other concerned persons are informed regarding this Field Safety Notice. 4. Complete the enclosed customer reply form and return it to NxStage. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices.
Initiated
2017-11-10
Root cause
Labeling design
Terminated
2019-12-06
Distribution
US nationwide distribution
Recall number
Z-1132-2018
Recall status
Terminated
CDRH record ID
161806
FDA record created
2018-03-19
Recall event number
79276

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗