Skip to content

Public FDA record

Z-1132-2021: LumiraDx SARS-CoV-2 Ag Test Strip Kit US EUA (48 Tests / EN). LumiraDx Severe Acute Respiratory Syndrome

Recall record Z-1132-2021 for LumiraDx. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1132-2021
Company in FDA record
LumiraDx
FDA classification date
Feb 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Observations of potential false positive results in the specified lots. Use of these test strips may result in false positive patient test results and potential exposure to unnecessary treatment or quarantine.

Source facts

Action
On 01/28/2021, Lumiradx issued Urgent Medical Device Recall notices to customers via letter notifying customers the potential of false positive results in specified lots of LumiraDx SARS-CoV-2 Antigen (Ag) Test Strip. A false positive test result of a patient could result in potential exposure to unnecessary treatment or quarantine. (Updated 3/15/2021) On 2/26/2021, Lumiradx issued a second Urgent Medical Device Recall notice to customers via letter expanding the recall to include 3 additional lots. These lots are subject to recent feedback from a small number of customers regarding a small to moderate elevation in numbers of reported false positive patient test results.
Initiated
2021-01-11
Root cause
Nonconforming Material/Component
Terminated
2021-12-03
Distribution
Worldwide distribution. US states of CA, GA, PA, SD. (Updated 3/15/2021) Austria, Germany, Ireland, Italy, Lebanon, Netherlands, Slovenia, Spain, United Kingdom.
Recall number
Z-1132-2021
Recall status
Terminated
CDRH record ID
185635
FDA classified
2021-02-17
Recall event number
87259

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗