Z-1132-2023: EVIS EXERA III Colonovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (in

Recall record Z-1132-2023 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1132-2023
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Feb 13, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination

Source facts

Action
Olympus issued Urgent Field Safety Notice communication for the customers without the device on December 21, 2022 via FedEx 2-day delivery.. Letter states reason for recall, health risk and requires the following action: Although the device is no longer in your possession, Olympus requires you to take the following action: 1. Evaluate the situation according to your hospital procedures to determine if additional actions, including patient notification and/or clinical follow up, may be required. 2. Review affected patient records and report any potentially related adverse events (such as infection) to Olympus and FDA (instructions below). 3. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. Go to https://olympusamerica.com/recall. Enter reference number 0419 and provide your contact information as indicated in the portal. Olympus requests that you report any complaints, including infections, to Olympus and the FDA. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Contact (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter. ***Update*** Firm provided recall delivered Urgent Medical Device Recall to one consignee dated 1/19/23 on 1/20/23 to customer. Letter states: Inspect your inventory for this model and serial number. Your local Sales Representative will assist in the return of the unit to Olympus and arrange either an exchange for a Certified Pre-Owned scope or a credit at fair market value. 2. Evaluate the situation according to your hospital procedures to determine if additional actions, including patient notification and/or clinical follow up, may be required. 3. Review affected patient records and report any potentially related adverse events (such as infection) to Olympus and FDA (instructions below). 4. Acknowledge receipt of this letter by e-mail response to your Sales Representative or to me directly a
Initiated
2022-12-21
Root cause
Process design
Distribution
US Nationwide distribution in the states of CA, FL, GA, NY, OR, VA.
Recall number
Z-1132-2023
Recall status
Open, Classified
CDRH record ID
197678
FDA classified
2023-02-13
Recall event number
91411

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗