Z-1133-2021: IOLMaster 700

Recall record Z-1133-2021 for Carl Zeiss Meditec AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1133-2021
Company in FDA record
Carl Zeiss Meditec AG
FDA record creation date
Feb 18, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen

Source facts

Action
On 12/11/2020, the firm sent a "FIELD SAFETY CORRECTION ACTION URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS informing them about a software error with software versions 1.90.2.09 and 1.90.8.06 that when used in conjunction with modality worklist functionality, is not performing as expected. These issues may lead to false display of patients in the patient manager screen, which can subsequently lead to the implantation of a wrong intraocular lens (IOL). The Recalling Firm is informing it customers that a software update needs to be installed on their device(s), or the modality worklist functionality (MWL) needs to be deactivated in case where they are unable to install the software update immediately. Software update will be available starting December 7th 2020 11 am ET via www.zeiss.com/iolmaster-700-software-update-us. The software will also be made available on a USB drive at a later point. If customers are using MWL, the Recalling Firm is asking them to repeat all measurements and IOL power calculations of patients that have underwent measurements with software 1.90.2.09 or 1.90.8.06 and have not yet undergone surgery to mitigate the risk of wrong patient measurement assignments and implantation of the incorrect IOL. Customers are asked to complete and return the Acknowledgement and Receipt Form via e-mail to USFCA.meditec.us@zeiss.com or by fax 925-448-3880 For any questions, contact Customer Support at 877-486-7473.
Initiated
2020-12-07
Root cause
Software design
Terminated
2022-06-24
Distribution
U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Austria, Australia, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, France, United Kingdom, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Japan, Kuwait, Malaysia, Netherlands, New Zealand, Philippines, Poland, Portugal, Russia, Sweden, Singapore, Slovakia, Thailand, Taiwan, and South Africa.
Recall number
Z-1133-2021
Recall status
Terminated
CDRH record ID
185137
FDA record created
2021-02-18
Recall event number
87055

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗