Skip to content

Public FDA record

Z-1134-2014: DeRoyal ST. Circumcision Clamp, REF 32-1622, Size 1.3 cm, 1 Per Pack, Rx Only STERILE EO, distributed by

Recall record Z-1134-2014 for DeRoyal Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1134-2014
Company in FDA record
DeRoyal Industries Inc
FDA classification date
Mar 4, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Three lots of circumcision clamps were manufactured out of specification and as a result may not properly clamp at the surgical site completely.

Source facts

Action
DeRoyal initiated the recall by sending overnight letters to its direct accounts on 01/16/2014. If distributors inform DeRoyal of further distribution, DeRoyal will contact those consignees in the same manner. If necessary, further communications will be sent in the form of another mailing, phone calls, and emails to materials management personnel when possible.
Initiated
2014-01-16
Root cause
Process control
Terminated
2015-10-13
Distribution
US Distribution to states of: NJ, AL, TX, WA, IL, OK, VA, FL, OR, and KS.
Recall number
Z-1134-2014
Recall status
Terminated
CDRH record ID
125456
FDA classified
2014-03-04
Recall event number
67442

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗