Skip to content

Public FDA record

Z-1134-2026: Brand Name: Ion Endoluminal System Product Name: Ion Vision Probe Bag Model/Catalog Number: 490127 Softwa

Recall record Z-1134-2026 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1134-2026
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Jan 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to probe bags that may have a compromised or incomplete sterile pouch seal.

Source facts

Action
On 12/19/2025, the firm emailed to customer an "Urgent Medical Device Recall" letter informing them that Intuitive has identified that some Vision Probe Bags may have a compromised or incomplete sterile pouch seal. Investigation determined that during the supplier manufacturing and packaging process, the Vision Probe Bag is placed into a separate sterile pouch for sealing. During this step, air trapped inside the Vision Probe Bag can remain within the pouch and expand during sterilization, which may place stress on the sterile-pouch seals and result in partial seal failure, creating a potential sterility breach. Customers are instructed: 1. Identify affected product: Inspect the product label and confirm the product number and lot number. a. Appendix A lists affected Part Numbers and Lot Numbers b. Appendix B provides an example label highlighting the location of the Part Number and Lot Number. 2. If the label displays lot numbers within the affected range shown in Appendix A, inspect the product pouch seal. Do not use the product if the seal is compromised or has an incomplete seal and scrap the affected units in accordance with your facility s disposal procedures. 3. Contact Intuitive Customer Service to obtain credit for any scrapped units. a. Product return to Intuitive is not required. 4. Retain a copy of this letter and the complete Acknowledgment Form for your records. 5. Ensure all relevant personnel within your facility are informed of this correction. For Questions - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com " India: +1-800-103-6952 (9 am to 6 pm IT) Taiwan: +0800-86-8181 (9 am to 6 pm CT)
Initiated
2025-12-04
Root cause
Process control
Distribution
U.S.: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MA, MD, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. O.U.S.: Australia, China, France, Germany, Italy, Japan, South Korea, Spain, Switzerland, and United Kingdom
Recall number
Z-1134-2026
Recall status
Open, Classified
CDRH record ID
217909
FDA classified
2026-01-21
Recall event number
98215

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗