Open original FDA record ↗Recall record
- FDA identifier
- Z-1135-2010
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Mar 15, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Paddle shock switch may fail to actuate and delay or prevent delivery of defibrillation therapy
Source facts
- Action
- Philips issued Urgent Medical Device Recall notification to customers by letter on 1/25/10 via UPS advising of this defect and providing instructions on how to determine if they have a defective paddle. The customers will be assisted by field service engineers to remove and replace these defective units (those within the date range of September 1, 2008 to December 30, 2008). Customers are asked to follow the Actions to be taken by customer/user section of the Urgent Medical Device Recall Notification. Removed units will be replaced free of charge and will be destroyed by Philips. Distribution of the Urgent Medical Device Recall Notification will be managed by Philips representatives in each affected geography. Customers are asked to follow the Actions to be taken by customer/user section of the Urgent Medical Device Recall Notification.
- Initiated
- 2010-01-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2021-03-16
- Distribution
- Worldwide distribution: USA, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, Estonia, Finland, France, Georgia, Germany, HONG KONG, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Lebanon, Lithuania, Malaysia, Mauritius, Netherlands, NICARAGUA, Norway, Peru, Philippines, POLAND, Portugal, Puerto Rico, Russian Federation, Saudi Arabia, Serbia & Montenegro, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad/Tobago, Turkey, UKRAINE, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, and Vietnam.
- Recall number
- Z-1135-2010
- Recall status
- Terminated
- CDRH record ID
- 88133
- FDA classified
- 2010-03-15
- Recall event number
- 54282