Open original FDA record ↗Recall record
- FDA identifier
- Z-1135-2012
- Company in FDA record
- Boston Scientific CRM Corp
- FDA classification date
- Mar 1, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
One (1) Model 6476 LATITUDE Communicator is unable to interrogate the patient's pulse generator. Boston Scientific has determined that this is due to a patient receiving an incorrect communicator. Therefore the communicator did not find and was unable to interrogate the assigned patients PG. Additionally when the patient attempted to set-up this incorrect communicator, the system reflected
Source facts
- Action
- On January 19, 2012 Boston Scientific telephoned the affected patient and their clinic. The patient was notified that the communicator they received will not be able to activate successfully and that their communicator and sensors would be replaced. The clinician was informed of the patient's communicator was sent incorrectly to the other patient who attempted to activate it, the monitored status of the website was incorrect and that a new communicator would be shipped to the patient that will work with their device.
- Initiated
- 2012-01-19
- Root cause
- Packaging process control
- Terminated
- 2012-07-05
- Distribution
- Nationwide distribution: UT only.
- Recall number
- Z-1135-2012
- Recall status
- Terminated
- CDRH record ID
- 107311
- FDA classified
- 2012-03-01
- Recall event number
- 61107