Skip to content

Public FDA record

Z-1136-2021: VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sa

Recall record Z-1136-2021 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1136-2021
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA record creation date
Feb 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A deformity affecting a specific lot of VITROS Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.

Source facts

Action
On 01/18/2021, Ortho Clinical Diagnostics issued an Urgent Product Correction Notification via e-mail due to a a deformity in a specific lot of VITROS Chemistry Products VersaTips. The affected product is to be no longer distributed and discarded.
Initiated
2021-01-14
Root cause
Under Investigation by firm
Terminated
2022-12-27
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada.
Recall number
Z-1136-2021
Recall status
Terminated
CDRH record ID
185527
FDA record created
2021-02-19
Recall event number
87200

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗