Open original FDA record ↗Recall record
- FDA identifier
- Z-1136-2021
- Company in FDA record
- Ortho-Clinical Diagnostics, Inc.
- FDA record creation date
- Feb 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A deformity affecting a specific lot of VITROS Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.
Source facts
- Action
- On 01/18/2021, Ortho Clinical Diagnostics issued an Urgent Product Correction Notification via e-mail due to a a deformity in a specific lot of VITROS Chemistry Products VersaTips. The affected product is to be no longer distributed and discarded.
- Initiated
- 2021-01-14
- Root cause
- Under Investigation by firm
- Terminated
- 2022-12-27
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada.
- Recall number
- Z-1136-2021
- Recall status
- Terminated
- CDRH record ID
- 185527
- FDA record created
- 2021-02-19
- Recall event number
- 87200