Open original FDA record ↗Recall record
- FDA identifier
- Z-1136-2025
- Company in FDA record
- Cardinal Health 200, LLC
- FDA classification date
- Feb 10, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
Source facts
- Action
- An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.
- Initiated
- 2024-11-08
- Root cause
- Material/Component Contamination
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.
- Recall number
- Z-1136-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211611
- FDA classified
- 2025-02-10
- Recall event number
- 95947