Z-1136-2026: TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN16

Recall record Z-1136-2026 for Sysmex America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1136-2026
Company in FDA record
Sysmex America, Inc.
FDA classification date
Jan 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.

Source facts

Action
A Product Notification Field Action was sent to customers on 12/15/25. Actions for customers 1. To prevent injury, never place your hands around the lift section at any time. 2. In addition, please follow these instructions when handling rack jams or similar errors: " Pay special attention not to insert your hands around the lift section during any intervention. " If a rack, tube, or any object becomes lodged around the lift section, turn off the power. After confirming that the lift section has moved to the bottom dead point, carefully remove the rack and blood collection tubes. " If the rack or blood collection tubes are jammed in the lift section and cannot be removed with one hand, contact Sysmex for assistance. 3. Sysmex will implement a software update to improve lift position control to prevent unintended sudden drops of the lift section by ensuring that, after an error occurs, the lift automatically returns to the home position. Our service team will contact you to schedule the installation of software version 00-18 by 12/31/2025. 4. Software update acknowledgement a. To assist Sysmex with the field safety communication, please acknowledge the communication by proceeding to the URL below to complete and submit the Acknowledgement Form. See Figure 1. URL: https://pages.sysmex.com/TS-10-Acknowledgement.html b. On the Acknowledgement Form at the URL, you will be directed to enter First/Last Name, Role/Title, Email Address, Phone #, Facility Name, Street Address, City, State / Province, Zip Code, Device Serial Number and acknowledge receipt of the notice by selecting Yes. c. Once the Acknowledgement Form is completed, a confirmation email will be sent to your email address. NOTE: If you do not see the email confirmation, check the SPAM or JUNK folders of your email box. 5. Please distribute this Product Notification as appropriate to your laboratory staff. 6. File this Product Notification as a part of your laboratory s Quality System as required. Questions
Initiated
2025-12-15
Root cause
Software design
Distribution
US Nationwide.
Recall number
Z-1136-2026
Recall status
Open, Classified
CDRH record ID
217906
FDA classified
2026-01-21
Recall event number
98213

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗