Z-1137-2019: Biomerieux Vitek 2 Test kit VITEK 2 is an automated system consisting of instruments, software and reag

Recall record Z-1137-2019 for bioMerieux, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1137-2019
Company in FDA record
bioMerieux, Inc.
FDA record creation date
Apr 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unexpected ESBL (Extended Spectrum Beta Lactamase) phenotype had been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N254, AST-N330, AST-N371 (ref.413723,418674, 422024) test kits.

Source facts

Action
On 03/15/2019, the firm, BioMerieux sent notifications to all their consignees with the following instructions: "Please take the following actions at this time: Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. Apply the VITEK 2 Systems bioART rule criteria as previously mentioned (i.e., using the criteria listed below for ESBL / E. coli isolates). Refer to Appendix A for detailed instructions. o If the card is AST-NXXX (where XXX indicates the card type designation, 254 / 330 / 371) o The organism is Escherichia coli o And the phenotype is ESBL o And the drug cefuroxime is <=8 o And the drug ceftazidime is <=0.5 o Then stop for consultation and add a comment: There is no evidence that the isolate is an ESBL producer. The category results for cephalosporins should be interpreted as tested, before the AES expertise. Please store this letter with your bioMrieux instrument documentation. Complete the Acknowledgement Form and return it to your local bioMrieux representative. bioMrieux, Inc. is committed to providing our customers with the highest quality products, and we apologize for any inconvenience this may have caused in your laboratory. If you have any questions or concerns, please contact your local bioMrieux representative."
Initiated
2019-03-15
Root cause
Software change control
Terminated
2021-08-17
Distribution
Internationally to: United Kingdom, Latvia, Poland, Ukraine, Germany, and Scotland.
Recall number
Z-1137-2019
Recall status
Terminated
CDRH record ID
171710
FDA record created
2019-04-12
Recall event number
82501

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗