Z-1137-2021: PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 - Product Usage: The ven

Recall record Z-1137-2021 for Respironics California, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1137-2021
Company in FDA record
Respironics California, LLC
FDA record creation date
Feb 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to unqualified "design verification" batteries being shipped to customers.

Source facts

Action
On 01/06/2021, the firm contacted via telephone calls and then on 01/29/2021 sent a follow-up "URGENT - Medical Device Removal" letter via priority mail to customers to inform them that the Recalling Firm shipped unqualified batteries to customers. These batteries are similar to production batteries but did complete final qualification testing. Potential hazards that may result, In a worst-case scenario, a battery failure could occur while in use on a patient during intra-hospital transport or when not connected to AC power. In both situations, the patient could lose ventilatory assistance and experience severe hypoxemia. Customers are provided instructions on how to identify affected products: Check the label of backup batteries (PN 1076374) received on or after November 25, 2020, or installed on affected units. If the battery has LOT M94663-P1, it is affected. Customers are instructed to: 1. If the batteries have been installed in ventilators, remove the ventilators from service until a field service representative or approved service provider schedules a visit and replaces the battery. 2. If waiting for a field service representative or approved service provider is not possible given their current census, contact Recalling Firm by email at HRC.Recall.Response@Philips.com to request a replacement battery that trained biomed can install. 3. If the affected battery has not been installed in a ventilator, quarantine the battery until a Recalling Firm representative can replace it. 4. Complete, sign, and return the Acknowledgment and Receipt Form at the end of this letter. For further information and questions - contact the local representatives or Customer Care Solutions Center (24/7) +1-800-722-9377
Initiated
2021-01-06
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution including in the states of FL, GA, IA, IA, IL. IL, MN, MO, NC, ND, OH, TX, TX and UT.
Recall number
Z-1137-2021
Recall status
Open, Classified
CDRH record ID
185799
FDA record created
2021-02-19
Recall event number
87314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗