Skip to content

Public FDA record

Z-1137-2023: NovaStar TS, Anti-Asphyxia Valve, Reusable, Material Numbers MP01576 (Size S), MP01577 (Size M), and MP01

Recall record Z-1137-2023 for Draeger Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1137-2023
Company in FDA record
Draeger Medical, Inc.
FDA classification date
Feb 24, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.

Source facts

Action
An Urgent Medical Device Correction notification letter dated January 2023 was sent to customers. Actions to be taken: Use of the mask is contraindicated for patients and other persons in close vicinity to patients using the masks, that have implanted devices that may be affected by magnets, including but not limited to: Pacemakers Implantable cardioverter defibrillators (ICD) Neurostimulators Magnetic metallic implants/electrodes/valves placed in upper limbs, torso, or higher (i.e., neck and head) Cerebral spinal fluid (CSF) shunts (e.g., ventriculoperitoneal (VP) shunt) Aneurysm clips Embolic coils Intracranial aneurysm intravascular flow disruption devices Metallic cranial plates, screws, burr hole covers, and bone substitute devices Metallic splinters in the eye Ocular implants (e.g., glaucoma implants, retinal implants) Certain contact lenses with metal Implants to restore hearing or balance that have an implanted magnet (such as cochlear implants, implanted bone conduction hearing devices, and auditory brainstem implants) Magnetic denture attachments Metallic gastrointestinal clips Metallic stents (e.g., aneurysm, coronary, tracheobronchial, biliary) Implantable ports and pumps (e.g., insulin pumps) Hypoglossal nerve stimulators Devices labeled as MR (Magnetic Resonance) unsafe Magnetic metallic implants not labeled for MR or not evaluated for safety in a magnetic field Apart from the contraindication, you can continue to use NovaStar masks, if you ensure that the magnets are kept at least 15 cm (approx. 6 inches) away from any other medical implants or medical devices that can be impacted by the magnetic fields. According to our records, your facility has received NovaStar TS masks material number MP01576, MP01577, MP01578, MP01579, MP01580, and/or MP01581, manufactured by Draegerwerk AG & Co. KGaA. Please add a copy of this Urgent Medical Device Correction Notification to every box of masks you have in your facilit
Initiated
2023-01-23
Root cause
Error in labeling
Distribution
US Nationwide Distribution
Recall number
Z-1137-2023
Recall status
Open, Classified
CDRH record ID
198130
FDA classified
2023-02-24
Recall event number
91538

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.