Open original FDA record ↗Recall record
- FDA identifier
- Z-1137-2026
- Company in FDA record
- Community Products, LLC
- FDA classification date
- Jan 22, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for fraying on the body support strap.
Source facts
- Action
- Community Products notified consignees via email on about 12/18/2025. Consignees were instructed to notify all customers, if further distributed, of the recall, inspect all affected devices for visible fraying, and cease use if any is observed. Consignees were requested to provide contact information for the individual responsible for performing the replacement. If devices were further distributed, consignees were advised to manage all customer outreach, inspection and replacement. The other option was to provide Community Products with customer contact information, and Rifton will manage the process. Lastly, consignees were requested to confirm receipt of the notification.
- Initiated
- 2025-12-18
- Root cause
- Process change control
- Distribution
- Worldwide distribution - US Nationwide and the countries of UAE, Australia, Canada, EU, Israel, Hong Kong, Japan, Korea, New Zealand, Russia, Saudi Arabia, and UK.
- Recall number
- Z-1137-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217700
- FDA classified
- 2026-01-22
- Recall event number
- 98140