The ultrasound system used in certain settings may result in overheating of the ultrasound probe head.
Source facts
Action
The January 3, 2018, notification letter requests the consignee complete and return a recall response form. The letter states that Esaote customer service will contact the consignee about the software update. In the meantime, it is recommended that the consignee does not use the PA023E probe in Pulsed Wave (PW) Doppler mode and that they refrain from using the system in Elaxto mode where the patient is unconscious or sedated. The consignee is asked to check the temperature of the probe before use.
Initiated
2018-01-03
Root cause
Software design
Terminated
2019-05-30
Distribution
US distribution in the states: AZ, FL, OK, PA, and TX.
Recall number
Z-1138-2018
Recall status
Terminated
CDRH record ID
162617
FDA record created
2018-03-20
Recall event number
79547