Z-1138-2019: IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in th

Recall record Z-1138-2019 for Topcon Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1138-2019
Company in FDA record
Topcon Medical Systems, Inc.
FDA classification date
Apr 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.

Source facts

Action
Topcon issued Device Correction Notice on 3/22/19 providing problem- there exists a situation in which the displayed patient name and the thumbnail image do not match and action to take.Topcon has corrected this issue in IMAGEnet 6 version 2.0.3 and will be providing a remote upgrade free of charge. Contact customer support team to schedule the upgrade at 866.922.6278 option 1 or by email at TMSservice@topcon.com
Initiated
2019-03-06
Root cause
Labeling design
Terminated
2019-07-17
Distribution
Nationwide
Recall number
Z-1138-2019
Recall status
Terminated
CDRH record ID
171349
FDA classified
2019-04-12
Recall event number
82322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗