Complaints received that device used to prepare femur to receive femoral implant would split during surgery.
Source facts
Action
Firm sent notification letter to consignees on 08/25/2006, along with replacement speedblocks. Consignees asked to trade out old speedblocks with the replacements.
Initiated
2006-08-15
Root cause
Other
Terminated
2008-03-27
Distribution
Product distributed to direct accounts in USA and one consignee in Japan.
Recall number
Z-1139-2007
Recall status
Terminated
CDRH record ID
52040
FDA classified
2007-08-07
Recall event number
37878
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.